    題名: A preliminary report of a disposable electrical non-fiberoptic endoscope in thoracoscopic surgery
    作者: Chen, CH (Chen, Chih-Hao)
    Chang, H (Chang, Ho)
    Yang, LY (Yang, Liang-Yu)
    Liu, HC (Liu, Hung-Chang)
    Tsung, TT (Tsung, Tsing-Tshih)
    Hung, TT (Hung, Tzu-Ti)
    貢獻者: Dept Mech Engn
    關鍵詞: Thoracoscopy/VATS
    Surgery
    ACCURACY
    CCD
    CMOS CAMERA
    日期: 2012
    上傳時間: 2013-02-25 14:48:21 (UTC+8)
    摘要: Purpose: Conventional thoracoscopic surgery requires a camera connected to optic fibers and rigid rod lens to ensure the provision of adequate light and quality of real-time images in the operative field. However, the camera, the connected optic fibers and rigid rod lens are not disposable due to cost, which is a concern as regards potential contamination of patients. To decrease such contamination, we designed a disposable device of extremely low cost which we tested in thoracoscopic surgery in animals.

    Description: A complementary metal-oxide-semiconductor is used for obtaining real-time image at a refresh rate of 30 frames per second. A circumferential light was added by a light emitting diode. We connected wires to a universal serial bus adapter, with which the device can negotiate with a computer so as to control signal retrieval and adjustment of the light as well as focus. The device was designed to be as compact as possible. The contour resembled a conventional thoracoscope, but with no optic fibers and rigid rod lens included.

    Evaluation: We used the devices to perform routine thoracoscopic surgical procedures, including wedge resection of the lung, lobectomy, esophagectomy, pericardiotomy and pleural biopsy in two 40-kg pigs under general anesthesia. The operating techniques were not altered while using this device.

    Conclusion: This disposable, electrical non-fiberoptic endoscope has the potential to be easily and safely used in routine thoracoscopic surgery at a minimal cost. Further clinical evaluation will be required to demonstrate the utility in human patients. (C) 2011 Surgical Associates Ltd. Published by Elsevier Ltd. All rights reserved.
    關聯: INTERNATIONAL JOURNAL OF SURGERY 卷: 10 期: 1 頁數: 20-24